تفاصيل المنتج
Features
1. Injection accuracy of up to ±2%
2. Anti-bolus and anti-siphon functionsincrease patient safety
3. Various visual and audible alarms
4. Automatically record historicalsetting parameters to save clinicaloperation time
5. Available change the infusionparameters without interruption ofinfusion
6. Keyboard locking function to preventmisoperation
7. Automatic identification of syringespecifications, suitable for different brand Syringe
8. Button and shuttle key double inputmode, to meet different users
9. Four infusion mode for choose
10. History record 50000 events
11. 12 months warranty
مواصفات
Syringe Size | 10, 20, 30, 50/60 ml |
KVO Rate | 0.1-1 ml/h (in 0.1 ml/h increments) |
Applicable Syringe | Compatible with syringe of any standard |
Drug Library | Available |
Weight&Size | 4.5 kg,245*120*115 m |
Anti-Bolus | Automatic |
Accuracy | ±2% (mechanical accuracy ≤1%) |
Occlusion Sensitivity | High, medium, low |
Safety Classification | Class Ⅱ, type CF |
Atmospheric Pressure | 700-1060 h |
Working Temperature | Temp:5-40℃,Relative Humidity 20-90% |
VTBI | 0.1-10000 ml ,<1000 ml in 0.1 ml increments, ≥1000 ml in 1 ml increments |
Additional Feature | Real-time infused volume, automatic power switching, automatic syringe identification, mute key, purge, bolus, anti-bolus, system memory |
Flow Rate | Syringe 10 ml: 0.1-300 ml/hSyringe 20 ml: 0.1-600 ml/h Syringe 30 ml: 0.1-800ml/hSyringe 50/60 ml: 0.1-1200 ml/ |
sInfusion mode | Memory mode Flow rate mode: ml/min, ml/h Time-based mode Body weight mode: mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h etc. |
Bolus Rate | Syringe 10 ml: 400ml/h Syringe 20 ml:600ml/h Syringe 30 ml: 900 ml/h Syringe 50/60 ml: 1300 ml/h |
Alarm | Occlusion, near empty, end program, low battery, endbattery,AC power off, motor malfunction, systemmalfunction, reminder alarm, pressure sensor error, syringe installation error,syringe drop off |
Battery | Battery life:7 hours at 5 ml/h Battery:9.6±1.6 V, rechargeable Power supply AC:110/230 V (optional), 50/60 Hz, 20 VA |